Statistical Tools and
Optoelectronic Measuring Instruments
Ionel Sabin and Ionel Ioana
Universitatea “Politehnica” Timişoara
Romania
1. Introduction
In the frame of European research projects, several air quality measuring campaigns in cross
roads, streets, parks as well in a non ecological waste deposit were realized. The analyzed
signals, representing CO, NO2, O3, SO2 and HC concentrations, were measured with several
optoelectronic instruments. Two of the utilized optoelectronic devices are shortly presented
at the beginning of the chapter.
Due to their random character, pollutant concentrations signals can be analysed using
statistical processing methods. The main statistical functions and parameters taken into
account within this chapter are histograms, correlation coefficients, correlation and
covariance functions (Ionel et al., 2009). Actually, statistical tools are usually utilized in
analysing ecological data (Zuur et al., 2007) but, as far we know, it is not common to imply
statistics in a comparative analysis of optoelectronic devices (Ionel et al., 2007).
Specific pre-processing procedures must be used for signal conditioning. Thus, „ideal“ low-
pass filtering based on fast Fourier transform can be implemented for the rejection of
measurement noise and artefacts from the pollutant concentration signals. On the other
hand, „ideal“ high-pass filtering allows the extraction of the variable component of the
pollutant level signals. In order to avoid redundant measurements, one can use
interpolation for increasing the number of samples, especially in the case of slowly varying
meteorological parameters.
Computer experiments with real pollutant concentration signals lead to some practical
recommendations concerning acquisition parameters like data size and sampling
frequency. The most important practical rules are as follow: assure the temporal length of
the measured signal, assure the necessary resolution on the time axis, and make
interactive verifications of the acquisition parameters during the measuring campaign.
Guidance on MATLAB software for calculating statistical functions an
d parameters are
provided.
As a particular application, the correlative comparison of two carbon monoxide (CO)
measuring instruments is presented. The point source device and the open path optical
remote sensing instrument do actually not measure the same quantity but a statistical
comparison of the two instruments is still possible. The correlative analysis leads to the
expected conclusion that the open path instrument is more suitable for monitoring the
pollution level in a large area than the classical point source device.
www.intechopen.com
412 Optoelectronic Devices and Properties
2. Pollutant concentrations measured with optoelectronic instruments
2.1 The optoelectronic measuring instruments
One of the utilized instruments was the specialized HORIBA APMA-350E CO monitor,
which furnishes the local pollution level. Fig. 1 presents a bloc diagram of this instrument
working on the classical Non-Disperse Infrared (NDIR) method. The APMA-350E
represents a generation of ambient CO monitors designed to eliminate routine calibration
cycles and to provide long-term stable measurements and unattended continuous
operation. It features a newly developed cross-flow modulation (CFM) technique which
results in remarkable improved zero drift performance and sensitivity. The cross-flow
modulated analyzer incorporates the basic design features of the conventional NDIR
analyzer.
The essential new element in this design, according to Fig. 1, is a rotary valve that
alternately directs the sample gas and a reference gas to the one cell of the analyzer. By this
method, the distinction between the sample and the reference optical path is eliminated and
each path alternately functions as a reference and a sample path. The requirement for an
optical chopper to modulate the detector output is thereby eliminated. In the cross-flow
analyzer design, sensitivity is inherently increased because the amount of IR (infrared)
energy absorbed and translated into the output signal is theoretically doubled for any
concentration at the given modulation frequency. In addition, the signal-to-noise ratio is
significantly better because the optical chopper which tends to introduce noise in the
conventional NDIR instrument is removed in this CFM design. In the CFM scheme, gas flow
rates and cell configuration can be selecting providing very smooth modulation. To
minimize interference, dual detector system employing a compensating detector located
behind the main detector is adopted in this instrument. The two detectors are charged in
such a way that response to the interference gas in the second detector is compared to that
of the measured gas. The signal from this detector is amplified and subtracted from the
main detector signal, in the electronic part of the analyzer.
![]() |
The second utilized instrument was an IR HAWK system from Siemens Environmental
Systems, with the schematic diagram presented in Fig. 2. This instrument is an IR DOAS
(Differential Optical Absorption Spectroscopy) apparatus, which can be configured to detect
several species of pollutants including carbon monoxide. The beam path can be up to 400m
and detection is typically better than 50ppb. The HAWK system works by measuring the
absorption of infrared radiation passing along the instrument beam path by the gas to be
measured. The system consists of a monitor, which contains the source and the detector
unit, and a reflector. The total path length is therefore twice the distance between monitor
and reflector. The source emits over a range of wavelengths and the beam is modulated after
generation. The beam is reflected back to the monitor where it is filtered at a wavelength
specific to gas of interest. The filtered beam is focused onto a detector which compares
filtered and unfiltered reflected light in order to measure the concentration of the target gas.
Open path techniques have an advantage over the point source detectors: the sample volume
is mach greater, the non-uniformity of the sample is eliminated and a more representative
value of the concentration to be measured is obtained. Under field conditions, the degree of
mixing is affected by the local environment, primarily, wind and thermal gradients.
Fig. 3 shows a typical relative setup for the HAWK and HORIBA analysers. One should
observe also the meteorological mast, which continuously sent data (15 minutes mean
values) to the general data acquisition system.
sumber : Hoffmann, J. & Quint, F. (2007). Signalverarbeitung mit MATLAB® und SIMULINK®.
Anwendungsorientierte Simulationen, Oldenbourg Verlag, ISBN 978-3-486-58427-1,
München



Tidak ada komentar:
Posting Komentar